Autosampler injecting the next sample while the previous run is not yet finished
Applies to
ALEXYS system with Clarity software.
Summary
Autosampler already injecting the next sample while the previous run is not yet finished.
Cause and remedy
It is one of the software configuration settings.
When configuring a system with an AS110 using the ALIAS control module in Clarity always make sure that the sequence mode is not active (it is by default: off). The sequence mode is a special mode of operation in which the autosampler is ‘master’ and can autonomously execute a sequence of runs stored in the AS embedded memory. In that case, it will autonomously inject samples (with a time interval defined by the ‘autosampler runtime’) regardless of the run time programmed in the Clarity method/sample sequence. To avoid this erratic behavior: always have the sequence mode inactive!