Select Page
Welcome to our Knowledge Base

My bio-inert HPLC system for carbohydrates show a rapid decrease in peak heights

You are here:
< All Topics

Applies to

Agilent 1260II bio-inert HPLC system with PAD of carbohydrates at high pH.

Summary

The peak heights of carbohydrate standards decrease, and are not reproducible.

Cause

Certain pieces of bio-inert tubing need to be replaced with PEEK tubing (see picture blue and red indications). The tubing material is not pH resistant.

Remedy

There are a few ways to circumvent this issue, depending on the requirements.

  1. Order the MP35N replacement tubings from Agilent, pn 5005-6800 Pump capillary kit for ECD. This is the best option.
  2. Replace the 2 pieces of tubing by PEEK tubing of 500 um ID. There is a pressure rating though, which is 200 bar. In case the HPLC system pressure is higher than 200 bar use option 1 or 3.
  3. Replace the 2 pieces of tubing by stainless steel tubing of 600 um ID, and passivate the stainless steel with 15% nitric acid prior to use. This is best done by connecting the tubings to a syringe with some silicone tubing adapter and rinse the nitric acid through for 20 min. Wash out the nitric acid prior to putting the tubings in place.
Was this article helpful?
0 out Of 5 Stars
5 Stars 0%
4 Stars 0%
3 Stars 0%
2 Stars 0%
1 Stars 0%
5
How can we improve this article?
How Can We Improve This Article?
Table of Contents