Welcome to our Knowledge Base
Why do I need a Faraday shielded oven?
Applies to
Electrochemical detection
Answer
Electrochemical detection is characterized by its extremely low limits of detection. In order to get the best possible performance all sources of noise should be minimized. Such a source of noise is electrical interference from the environment. Also, temperature fluctuations will introduce noise and drift in baseline. All Antec’s detectors are equipped with a Faraday shielded oven compartment for placing the cell and HPLC column. It is an adequate remedy for electrical interferences and provides temperature stability.