ROXY installation

You are here:
< All Topics

Applies to

ROXY Exceed potentiostat with µ-PrepCell.

Recording

The video shows how to connect a syringe pump, and ROXY Exceed with a µ-PrepCell to an MS. The cell is filled completely with solvent, then the REF is placed. This is important to avoid trapping air in the REF connection fitting.

Before connection with an MS, a grounding T is placed for safety. After installation, a solution with insulin is supplied to the MS. This results in a steady signal at m/z 957 (insulin) until the cell is switched on. With the cell on, the reduction reaction starts and the disulfide bond that keeps the insulin chain A and B together is broken. Then the signal of Insulin drops, and the signal of m/z 681 and 851 appears, which is one of the insulin chains (reaction product).

 