Welcome to our Knowledge Base
ALEXYS system with poor reproducibility
Applies to
ALEXYS system, AS 100 autosampler
Summary
Poor reproducibility in peak height and peak area. Typically %RSD improves with concentration, however sometimes the %RSD is much worse than expected.
Cause
Possible causes:
- Autosampler
- most often: air in sampling tubing or syringe
- leakage of injection loop, valve, syringe, needle or tubing
- not enough sample in vial
- air bubble on bottom of micro vial
- contamination between injections
- Data acquisition
- baseline not correctly drawn
- unstable baseline, large noise, small peaks
- interfering peaks, shouldering peaks
- Pump
- leakage at connectors or seals
- Detector, HPLC
- contamination of working electrode
- reference electrode contaminated, chloride concentration not constant
- poor column performance, contamination of column
- not optimised detector settings
- carbohydrate analysis: carbonate in mobile phase increases ionic strength, constant shift in retention times
Remedy
- Autosampler
- remove air from tubing and syringe
- apply maintenance to autosampler
- needle wash ‘after each injection’ (in stead of ‘after each vial’)
- Data acquisition
- optimise acquisition parameters
- Pump
- apply maintenance, replace seals in case of leakage
- Detector, HPLC
- apply maintenance to flow cell
- add constant chloride concentration to REF or mobile phase (ISAAC)
- replace column (apply cleaning procedure first)
- optimise detector settings
- carbohydrate analysis: proper degassing and application of He to prevent carbonate from dissolving in mobile phase.