Seleccionar página
Welcome to our Knowledge Base

Why do I need a Faraday shielded oven?

You are here:
< All Topics

Applies to

Electrochemical detection

Answer

Electrochemical detection is characterized by its extremely low limits of detection. In order to get the best possible performance all sources of noise should be minimized. Such a source of noise is electrical interference from the environment. Also, temperature fluctuations will introduce noise and drift in baseline. All Antec’s detectors are equipped with a Faraday shielded oven compartment for placing the cell and HPLC column. It is an adequate remedy for electrical interferences and provides temperature stability.

Was this article helpful?
0 out Of 5 Stars
5 Stars 0%
4 Stars 0%
3 Stars 0%
2 Stars 0%
1 Stars 0%
How can we improve this article?
How Can We Improve This Article?
Table of Contents