Seite wählen
Welcome to our Knowledge Base

Excessive peak broadening with partial loop fill

You are here:
< All Topics

Applies to

Autosampler AS 100/110

Summary

Excessive peak broadening using partial loopfill for small samples.

Cause

Tubing connection syringe/needle exchanged. On the left the correct connection scheme for partial loopfill is shown. After loading the loop with a few microliters, the sample is injected in reversed flow onto the column. Using the scheme on the right results in peak broadening because the sample is diluted in the loop before entering the column.

Fig. 1. The correct connection scheme is shown on the left.

Diagnosis and Remedy

Connect to correct ports.

Was this article helpful?
4.5 out Of 5 Stars

1 rating

5 Stars 0%
4 Stars 100%
3 Stars 0%
2 Stars 0%
1 Stars 0%
How can we improve this article?
How Can We Improve This Article?
Table of Contents